10.0 ERRORMESSAGESANDTROUBLE‐SHOOTING
Error
Messages
Possible Causes Suggestions
Err E
Thememoryhasfailed.
Turnoffpowerswitchattherearofthescaleand
waitfortenseconds.Turnonagain.
Err 1
A/Dvalueislessthannormal
range,theADCortheloadcell
mayhavefailed.
Unplugtheloadcellconnection.Iftheproblemstill
existsthentheADCisfaulty.Ifnotthencheck
cablesandloadcells.
Err 2
A/Dvalueisbiggerthannormal
range,theADCortheloadcell
mayhavefailed.
Unplugtheloadcellconnection.Iftheproblemstill
existsthentheADCisfaulty.Ifnotthencheck
cablesandloadcells.
Err 3
Zeropointisabnormalwhen
poweron.
CheckADCcountsinparametersforstabilityor
verifyconnectionstoscale.
Err 4
A/Dvalueistoosmallwhen
calibratezeropoint.
Ensureloadcellisnotdamagedandthatanydead
loadisapplied.
Err 5
Rangecalibrationhasfailed.
Checkthatloadcelliswiredcorrectlyandisthe
correctsizeandoutputforscale.Ensurethat
correctmasswasloaded.
Err 6
Weighingvaluecannotbe
stabilizedincalibrationperiod.
Checkloadcellforliquidingress,orenvironmental
influences.Checkbatterycharge.
OUEr
Weighing value is bigger then
themaximumweighing.
Enter A/D value display, then check whetherthe
ADCisworkingcorrectly.Ensurethattheloadcell
hasnotbeenoverloaded.Checkparametersetting
formaxvalueorrecalibraterange.
Weighingdisplaywillnotshow
stablevalue.
Checkthebatteryforcorrectcharge.
Plugindifferentscaletotestifproblemiswiththe
scaleorindicator.
Ensureincrementvaluesettingisnottoosmall.
Cannotturnonthepower Ensurethatpowerswitchattherearisturnedon.
Plugin theadapterand ch eckthatthe“POWER”
LEDonthefrontoftheindicatorislit.Testvoltage
onthebatterytoensurethatitischargingabove6
volts.CheckloadcellandRS232connectiontosee
ifanyareshorted.
24|Page©AdamEquipmentCompany2008