6.3 CALIBRATION
The next section is the Calibration section.
A weight value and. “WEl” will be displayed.
• When WEI is displayed, press the [Unit] key to scroll through the
options for selecting the calibration mass.
• To select the desired mass press the [Mode] key.
• The raw ADC count will be displayed along with “CAI”.
• To set the zero calibration, press the [Mode] key again.
• The scale will display “0 LOD”, place the calibration mass
selected onto the top pan, the display will show the raw ADC
counts of the selected calibration mass, then press [Mode] again.
• The display shows PASS and then returns to normal weighing
• Remove the weight and continue to use.