press. Press <<Exit>>or <<Enter>> to return to weighing mode. If a programmable data
needs to be modified, press keys in the order of <<Modify>><<Figures>><<Enter>> to
enter new data. Data order is displayed as follows:
Pressing times Reminding symbols Data meaning
1 YXX printing is not allowed in weighing mode of
division=0.1e
2 A/D Weight A/D internal code
3 P09 Display the weight in zero point
4 P10 Display zero point upon powered on
5 P11 Calibrate zero point internal code
6 F Full Scale
7 E Check division 0.01~50, 12 options in total
8 P14 Displaying ratio
9 P15 Displaying Non-linearity edited value within
the scope of –1%Max~1%Max, weight unit
10 P16 Initial zeroing range 2~100 unit: %Max.
This parameter determined the difference
between present load and calibrated zero
point with 20%Max as default in first
zero-setting upon powered on.
11 P17 Zero-tracing switch, 0 is off and 1 is on
12 P18 Weight unit:
0=t 1=kg 2=g 3=mg
13 P19 bit number
0=600 1=1200 2=2400 3=4800 4=9600BPS
14 P20 Communication mode:
0= constant sending 1= answer style
15 P21 Communication address:
1~255
16 P22 Printer styles:
0=without Chinese characters database
1=with Chinese characters database (only
the printer with Chinese characters
database is applicable at present, this para.
does not work)
17 P23 The lighting time of the background light
1~255, unit: Second. 0=no automatic off
18 P24 Scoreboard switch:
0=on 1=off
19 P25 RS232 switch
0= on 1=off
20 P26 Displaying divisions 0.01~50 12 options
total
21 P27 Displaying Version No. For example: H2C