NI PXI-5600 Calibration Procedure 8 ni.com
Calibration Flowchart
Figure 2 shows the main steps in the NI PXI-5600 calibration process.
Figure 2. Calibration Process Flowchart
No
Finished.
Starting with temperature <30 °C,
verify temperature correction coefficient.
Wait 20 minutes for the
NI PXI-5600 to warm up.
Verify internal reference
frequency accuracy.
Verify absolute and relative
amplitude accuracy.
Ye s
Power down the system, wait for
the NI PXI-5600 to cool down, then restart.
Starting with temperature <30 °C,
adjust temperature correction coefficient.
Wait 20 minutes for
the NI PXI-5600 to warm up.
Adjust internal reference frequency.
Adjust IF frequency response.
Adjust RF frequency response.
Repeat verification steps after adjusting.
Any results out
of specification?