Instrument Location Requirements
To ensure reliable read switch activation, the instrument must sit on a reasonably flat, level,
and stable surface.
Strip Feeding Requirements
To ensure consistent strip readings, you must feed the strip into the instrument at a
continual pace (without pausing) and speed before it is taken in by the drive motor. More
specifically, a strip cannot be fed into the instrument too quickly (more than 7.2
inches/second or 183 mm/second), too slowly (less than 2.6 inches/second or 66
mm/second), or intermittently (pausing one or more times). If you are experiencing any
strip-feeding problems, observe how fast the motor is moving the strip (about 5.4
inches/second or 137 mm/second) and try to match the speed as closely as possible.
Before feeding a second strip or a second pass of a multi-pass strip into the instrument, you
must momentarily pause (approx. 3 seconds) to allow the instrument to process data from
the first measurement. The LED is not lit when the instrument is processing data.
Instrument LED Indicator
The LED indicates a variety of instrument operation conditions, such as calibration mode
and operation. Below is a complete list of conditions reported by the LED.
Operation Mode
Solid Green—self-test passed and instrument is ready for use with defined strip.
Slow Flashing Green—strip reading in progress (strip being read) or ready for next pass
(multi-pass strip definition).
Fast Flashing Green—strip reading was attempted but the wrong strip was read, or strip
was skewed during reading.
Slow Flashing Amber/Green—a configuration barcode was scanned and the instrument is
waiting for the 2
nd
or 3
rd
barcode pass.
Fast Flashing Amber/Green—there was an error in recognizing a barcode or the
configuration failed. Reread the entire barcode, starting with the first pass.
Calibration Mode
Slow Flashing Green—instrument calibration in progress (strip being read).
Solid Green after Flashing Green—instrument calibration was successful and instrument
is ready for use.
Fast Flashing Green—instrument calibration failed.
Error/Reset Mode
NOTE: When the instrument is first turned on, the motor runs, and the LED illuminates red while
performing self-tests. Self-tests are completed in approximately 2 to 3 seconds, at which time the LED
turns green and the motor stops under normal conditions.
Fast Flashing Red—incorrect power supply attached. Verify correct 12-volt power supply
is used.
Slow Flashing Red—a factory reset is in progress.
Flashing Red/Green— the host firmware loader program is reprogramming the
instrument.