Section 3 - Analyzer Setup
3.3 Temperature Calibration (Cont.)
5.
TEMP CAL 160C
CURRENT TEMP: 100 C
TM TO CAL PNT: 15:00 MIN
ADJ CAL READING: 160C
ACCEPT NEW CAL
ABORT CALIBRATION
TEMP CAL 160C
CURRENT TEMP: 160 C
TM TO CAL PNT: 10:00 MIN
ADJ CAL READING: 160C
ACCEPT NEW CAL
ABORT CALIBRATION
ADJUST TO THERMOMETER READING
START CALIBRATION
160C
TEMP CAL 160C
CURRENT TEMP: 160 C
TM TO CAL PNT: 10:00 MIN
ADJ CAL READING: 161 C
ACCEPT NEW CAL
ABORT CALIBRATION
TARE
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
3.4 Time and Date Setting
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
1.
ANALYZER SETUP
WEIGHT CAL
TEMP CAL
TIME-DATE
RS-232
PRINT
EXIT
TO START
TIME - DATE
SET FORMAT
FORMAT: MM/DD/YR
SET DATE: 07/10/01
TIME FMT: 12HR
SET TIME: 11:30 AM
2.
TIME - DATE
FORMAT: DD.MM.YR
SET DATE: 10. 07. 01
TIME FMT: 12HR
SET TIME: 11:30 AM
15 minute time lapse
7.
Open the cover, allow unit to cool, remove the
calibration fixture, replace the pan support and pan.
TARE
TARE
AFTER 15 MINUTE TIME PERIOD
6.
TEMP CAL 160C
CURRENT TEMP: 160 C
TM TO CAL PNT: 10:00 MIN
ADJ CAL READING: 161 C
ACCEPT NEW CAL
ABORT CALIBRATION
TARE
8.
ACCEPT CALIBRATION
FINISH CALIBRATION
or
x2
or