A SERVICE OF

logo

Parameter Settings Transfer Messages
Setting up your Monitor 3-63
Setting up your
Monitor
Parameter Settings Transfer Messages
The following messages related to Parameter Settings Transfer can
appear on the display:
Message Condition Action required
(Parameter name)
no settings
transfer capability
The indicated parameter
module is not a T
module.
Insert a module that has
a T on the front.
(Parameter name)
check settings''
No settings are stored in
the module, or the
settings are incorrectly
stored.
Enter the parameter
Task Window and check
the settings. The
settings are
automatically stored
when you exit the Task
Window.
Check Patient
Category and
(parameter name)
settings
The patient category
stored with the settings
in the module is
different from that of the
monitor.
Either change the
patient category for the
monitor or change the
parameter settings.
(Parameter name)
unable to store
settings in module
The parameter module
was removed before the
settings could be stored,
or there is a malfunction
in the module. (This
message can appear with
Settings Transfer On or
Off.)
Press ,
re-insert the module and
check/adjust the
settings in the
parameter Task
Window. Exit the Task
Window and leave the
module in the rack for
about 10 seconds before
you remove it. If the
same message appears,
replace the module.
Silence Reset