© National Instruments Corporation 13 SCXI-1125 Calibration Procedure
Specifications
The following table contains the specifications you should use when
verifying and calibrating your SCXI-1125 module. If your device has been
calibrated within the last year, the test point value should fall between the
Upper Limit and Lower Limit values.
Table 1. SCXI-1125 Specifications at 25 °C ±10 °C
Gain Range Test Point Upper Limit (V) Lower Limit (V)
1 5 4.950000 4.9623181 4.9376819
1 0 0.000000 0.005002 0.005002
1 5 4.950000 4.9376819 4.9623181
2 2.5 2.475000 2.4811601 2.4688400
2 0 0.000000 0.002502 0.002502
2 2.5 2.475000 2.4688400 2.4811601
5 1 0.990000 0.9924652 0.9875348
5 0 0.000000 0.001002 0.001002
5 1 0.990000 0.9875348 0.9924652
10 0.5 0.495000 0.4962336 0.4937664
10 0 0.000000 0.000502 0.000502
10 0.5 0.495000 0.4937664 0.4962336
20 0.25 0.247500 0.24811781 0.24688220
20 0 0.000000 0.000252 0.000252
20 0.25 0.247500 0.24688220 0.24811781
50 0.1 0.099000 0.09924832 0.09875168
50 0 0.000000 0.000102 0.000102
50 0.1 0.099000 0.09875168 0.09924832
100 0.05 0.0495000 0.04962516 0.04937484
100 0 0.000000 0.000052 0.000052
100 0.05 0.049500 0.04937484 0.04962516
200 0.025 0.024750 0.02481358 0.02468642
200 0 0.000000 0.000027 0.000027