National Instruments SCXI -1125 Car Stereo System User Manual


 
SCXI-1125 Calibration Procedure 14 ni.com
200 0.025 0.024750 0.02468642 0.02481358
250 0.02 0.019800 0.01985126 0.01974874
250 0 0.000000 0.000022 0.000022
250 0.02 0.019800 0.01974874 0.01985126
500 0.01 0.009900 0.00992663 0.00987337
500 0 0.000000 0.000012 0.000012
500 0.01 0.009900 0.00987337 0.00992663
1000 0.005 0.00495 0.00496432 0.00493568
1000 0 0.000000 0.000007 0.000007
1000 0.005 0.004950 0.00493568 0.00496432
2000 0.0025 0.002475 0.00248316 0.00246684
2000 0 0.000000 0.000005 0.000005
2000 0.0025 0.002475 0.00246684 0.00248316
Limits Valid for 4 Hz or 10 kHz Filter Settings
Valid when using an MIO device after performing an MIO internal calibration.
Valid for both filter settings with 5.000 point averaging at 10,000 samples per second.
Table 1. SCXI-1125 Specifications at 25 °C ±10 °C (Continued)
Gain Range Test Point Upper Limit (V) Lower Limit (V)