A SERVICE OF

logo

Adjustment Procedures
5-60
2715 Spectrum Analyzer Service Manual
Signal Generator
Power Meter
50 Cable
50 Power
Sensor
50 Power
Sensor
Power
Divider
6dB
Pad (N)
20 dB
Pad (N)
Figure 5- 25: Test Equipment Characterization Using a Dual Channel Power Meter
for split50
Signal Generator
Power Meter
50 Cable
50 Power
Sensor
50
Termination
Power
Divider
6dB
Pad (N)
20 dB
Pad (N)
Figure 5- 26: Test Equipment Characterization Using a Single Channel Power Meter
to Characterize the Power Meter Output for split50
Signal Generator
Power Meter
50 Cable
50
Termination
50 Power
Sensor
Power
Divider
6dB
Pad (N)
20 dB
Pad (N)
Figure 5- 27: Test Equipment Characterization Using a Single Channel Power Meter
to Characterize the DUT Output f or split50