5. Advanced operation
Staff Ofce Corresponds to light conditions in ofce rooms.
Clinical Viewing Room Corresponds to light conditions in diagnostic
reading rooms for clinical viewing.
Emergency Room
Corresponds to light conditions in emergency
rooms.
Operating Room Corresponds to light conditions in operating rooms.
This setting has the highest maximum ambient light.
To select a reading room:
1. Bring up the OSD main menu.
2. Navigate to the Conguration > Calibration > Ambient Light menu.
3. Enter the Reading Room submenu.
4. Select one of the available reading rooms and conrm.
5.20 Continuous ALC
About Continuous ALC
Continuous ALC can only be selected when the DICOM display function is selected.
Therefore, please refer to "Display functions", page 25 to correctly set the display func-
tion.
Enabling continuous ALC will continuously recalculate the DICOM display function taking the averaged
ambient light into account.
To select continuous ALC:
1. Bring up the OSD main menu.
2. Navigate to the Conguration > Calibration > Ambient Light menu.
3. Enter the Continuous ALC submenu.
4. Select Enabled/Disabled as desired and conrm.
5.21 Image scaling
About image scaling
Enabling image scaling will multiply each individual pixel to one or more adjacent pixels so that the size
of the displayed image will be a multiple o f the original i mage source video input signal.
Image scaling is only possible when the resolution of your display’s video input signal
is less than or equal to half the maximum resolution of the display.
To enable/disable image scaling:
1. Bring up the OSD main menu.
2. Navigate to the Conguration > Image Source menu.
3. Enter the Scaling submenu.
4. Select Enabled/Disabled as desired and conrm.
K5902055 CORONIS FUSION 6MP 28/08/2013