Barco MDCC-6230 Car Video System User Manual


 
8. Important information
Immunity test
IEC 60601
Test levels
Compliance level
Electromagnetic
environment
guidance
Power frequency (50/60
Hz) magnetic eld
IEC 61000-4-8
3A/m
Not applicable
Power frequency
magnetic elds should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3Vrms
150 kHz to 80 MHz
3V/m
80 MHz to 2.5 GHz
3V
3V/m
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the Coronis Fusion
6MP, including cables,
than the recommended
separation distance
calculated from the
equation applicable
to the frequency
of the transmitter.
Recommended
separation distance
d=1.2P
d=1.2P 80 MHz to 800
MHz
d=2.3P 800 MHz to 2.5
Ghz
Where P is the maximum
output power rating
of the transmitter in
watts (W) according
to the transmitter
manufacturer and d
is the recommended
separation distance in
meters (m).
Field strengths from
xed RF transmitters,
as determined by an
electromagnetic site
survey,
2
should be less
than the compliance
level in each frequency
range.
3
Interference may occur in
the vicinity of equipment
marked with symbol:
K5902055 CORONIS FUSION 6MP 28/08/2013 43