Barco MDCC-6230 Car Video System User Manual


 
8. Important information
Immunity test
IEC 60601
Test levels
Compliance level
Electromagnetic
environment
guidance
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is af-
fected by absorption and reection from structures, obje cts and people.
Recommended separation distance
The Coronis Fusion 6MP is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer of the user of the Coronis Fusion 6 MP can help prevent elec-
tromagnetic interference by maintaining a minimum distance between p ortable and mobile RF communi-
cations equipment (transmitters) and the Coronis Fusion 6MP as recommended below, according to the
maximum output power of the c ommunications equipment.
Separation distance according to frequency of t ransmitter
Rated ma ximum output
power of transmitter
4
W
150kHz to 80MHz
d=1.2P
80MHz to 800MHz
d=1.2P
800MHz to 2.5GHz
d=2.3P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100121223
At 80 MHz and 800 MHz, the separation distance for the h igher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is af-
fected by absorption an d reection f orm structures, object and people.
2. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the Coronis Fusion 6MP is used exceeds the applicable RF compliance level above, the Coronis Fusion 6MP
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
re-orienting or relocating the Coronis Fusion 6MP.
3. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than3V/m.
4. For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter. Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
44 K5902055 CORONIS FUSION 6MP 28/08/2013