Fluke 5725A Car Amplifier User Manual


 
5725A
Instruction Manual
1-16
Wideband AC Voltage (Option -03)
Specifications apply to the end of the cable and 50 termination used for calibration:
Range
Absolute Uncertainty
±5°C from calibration temperature
30 Hz-500 kHz
Volts dBm
Resolution
24 Hours 90 Days 180 Days 1 Year
± (% output + µV)
1.1 mV -46 10 nV .4 + .4 .5 + .4 .6 + .4 .8 + 2
3 mV -37 10 nV .4 + 1 .45 + 1 .5 + 1 .7 + 3
11 mV -26 100 nV .2 + 4 .35 + 4 .5 + 4 .7 + 8
33 mV -17 100 nV .2 + 10 .3 + 10 .45 + 10 .6 + 16
110 mV -6.2 1
µ
V .2 + 40 .3 + 40 .45 + 40 .6 + 40
330 mV +3.4 1
µ
V 0.2 + 100 0.25 + 100 .35 + 100 .5 + 100
1.1V +14 10
µ
V .2 + 400 .25 + 400 .35 + 400 .5 + 400
3.5V +24 10
µ
V 15 + 500 .2 + 500 .3 + 500 .4 + 500
Frequency
Frequency
Resolution
Amplitude Flatness, 1 kHz
Reference
Voltage Range
Temperature
Coefficient
Settling
Time To
Full
Accuracy
Harmonic
Distortion
1.1 mV 3 mV >3 mV
Hz Hz ±% ±ppm/°C Seconds dB
10-30 .01 .3 .3 .3 100 7 -40
30-120 .01 .1 .1 .1 100 7 -40
120-1.2k .1 .1 .1 .1 100 5 -40
1.2k-12k 1 .1 .1 .1 100 5 -40
12k-120k 10 .1 .1 .1 100 5 -40
120k-1.2M 100 .2 + 3
µ
V.1 + 3
µ
V.1 + 3
µ
V 100 5 -40
1.2M-2M 100k .2 + 3
µ
V.1 + 3
µ
V.1 + 3
µ
V 100 0.5 -40
2M-10M 100k .4 + 3
µ
V.3 + 3
µ
V.2 + 3
µ
V 100 0.5 -40
10M-20M 1M .6 + 3
µ
V.5 + 3
µ
V.4 + 3
µ
V 150 0.5 -34
20M-30M 1M 1.5 + 15
µ
V 1.5 + 3
µ
V1 + 3
µ
V 300 0.5 -34
Additional Operating Information:
dBm reference = 50
Range boundaries are at voltage points, dBm levels are approximate.
dBm = 10 log (
Power
1 mW
),
0.22361V across 50 = 1 mW or 0 dBm
Minimum output: 300 µV (-57 dBm)
Frequency uncertainty: ±0.01%
Frequency resolution: 11,999 counts to 1.1999 MHz, 119 counts to 30 MHz.
Overload protection: A short circuit on the wideband output will not result in damage. After settling time,
normal operation is restored upon removal.