Fluke 5725A Car Amplifier User Manual


 
Introduction and Specifications
Specifications
1
1-17
General Specifications:
Warm-Up time: 2 X the time since last warmed up, to a maximum of 30 minutes.
System installation: Rear output configuration and rack- mount kit available.
Standard interfaces: IEEE-488, RS-232, 5725A, 5205A or 5215A, 5220A, phase lock in (BNC), phase
reference out (BNC).
Temperature performance: Operating: 0°C to 50°C.
Calibration: 15°C to 35°C. Storage: -40°C to 75°C.
Relative humidity: Operating: <80% to 30°C, <70% to 40°C, <40% to 50°C.
Storage: <95%, non-condensing.
Safety: Designed to comply with UL1244 (1987); IEC 348-1978; IEC 66E (CO) 4; CSA 556B.
Guard isolation: 20 volts
EMI/RFI: Designed to comply with FCC Rules Part 15, Subpart J, Class B; VDE 0871, Class B.
Reliability: MIL-T-28800D, para. 3.13.3.
Line Power: 47 to 63 Hz; ±10% allowed about selectable nominal line voltage: 100V, 110V, 115V, 120V,
200V, 220V, 230V, 240V. Maximum power: 5700A, 300 VA; 5725A, 750 VA.
Electromagnetic Compatibility: This instrument is designed to operate in Standards Laboratory
environments where in electromagnetic is highly controlled. If used in areas with RF fields >0.78 V/m there
could be errors in measurements.
Size:
5700A: Height 17.8 cm (7 in), standard rack increment, plus 1.5 cm (0.6 in) for feet; Width 43.2 cm (17 in),
standard rack width; Depth 63.0 cm (24.8 in), overall; 57.8 cm (22.7 in), rack depth.
5725A: Height 13.3 cm, (5.25 in); Width and depth same as 5700A. Both units project 5.1 cm, (2 in) from
rack front.
Weight: 5700A: 27 kg (62 lbs); 5725A: 32 kg (70 lbs).
43.2 cm (17 in)
17.8 cm (7 in)
63 cm (24.8 in)
6.35 cm (2.5 in)
FOR CABLE
ACCESS