CMODe Subsystem (TRIGger)
AM700 Audio Measurement Set Programmer Manual
3–49
CMODe Subsystem (Trigger)
The CMODe commands found here are used to control the Trigger modes of the
Audio Monitor.
CMODe:TRIGger:LEVel <numeric_value>
Usage: Sets the trigger level for the Audio Monitor trigger system.
Range: –173.62 to +173.62 V
Units: V or mV. Units follows the setting of the UNIT:VOLT command.
Query: CMOD:TRIG:LEV? returns the current trigger level setting in volts or
millivolts.
*RST: Sets trigger level to 0.00 V.
CMODe:TRIGger:MODE AUTO|NORMal
Usage: Select Auto or normal trigger mode for the Audio Monitor trigger system. Auto
trigger mode permits the acquisition to free run in the absence of a qualified
trigger signal and trigger when a qualified trigger signal is received. Normal
trigger mode requires that a qualified trigger signal of the correct level and
slope be received to trigger an acquisition. In normal mode the trace display is
erased in the absence of a trigger event.
Query: CMOD:TRIG:MODE? retuns the current trigger mode setting.
*RST: Sets the trigger mode to AUTO.