AMEasure Subsystem (Audio Analyzer)
AM700 Audio Measurement Set Programmer Manual
3–9
AMEasure[1–4]:REFerence:FILE:STORe
Usage: Command only. Stores the reference trace to the file name that was previously
set by the AMEasure[1–4]:REFerence:FILE[:NAME] command.
AMEasure[1–4]:REFerence:SET
Usage: Command only used to store or save a reference trace for comparison. The front
trace of the designated AME is copied for use as a reference. If the frontmost
trace is empty, it clears the reference memory. With regards to the measure-
ment, this is the same as pressing the “Set Ref” button on the front panel. The
“reference available” message should appear in the view window when the
reference is set.
AMEasure[1–4]:REFerence:STATe?
Usage: Query only, It returns 1 if, and only if, there is a valid reference for the given
measurement type.
Query: AME[1–4]:REF:STAT? returns 0 for no valid reference or 1 if there is a
valid reference for on for the designated AME.
AMEasure[1–4]:XY:TRACkgen <Boolean>
Usage: Sets or queries the state of TRACkgen for the designated AME in XY mode.
Example: AME1:XY:TRAC ON enables tracking of the generator
Parameters: ON|OFF or 1|0
Query: AME[1–4]:XY:TRAC? returns 0 for tracking disabled or 1 for enabled.
*RST: Sets XY mode Track Generator to tracking enabled.