A SERVICE OF

logo

Getting Started
AM700 Audio Measurement Set Programmer Manual
1–27
The FFT Analyzer application also contains the Multitone measurement calculate
blocks. A special SCPI command switches the FFT Analyzer between the two
functions as shown in Figure 1–12. The calculate block results are given different
trace names. The trace names for the FFT calculate block results are found using the
TRACe:CAT? command. The reply to the query is application specific and returns
the names for the Audio Analyzer traces when that application is running.
CMODe:FFT FFT|MTONe
CALCulate FFT
CALCulate
MULTITONE
CALCulate FFT
CALCulate
MULTITONE
Result1
Result2
Result3
Result4
Result5
Result6
Result7
Result8
Result9
Result10
Result11
Result12
Result13
Result14
Result15
Result16
DISPlay
Measurement
Input 1
Measurement
Input 2
CSTReam1
CSTReam2
TRACe:CAT?
TRACe:DATA? FFTMAG_1
Figure 1–12: Switching Between FFT and Multitone