AMEasure Subsystem (Audio Analyzer)
3–24
AM700 Audio Measurement Set Programmer Manual
NOTE. Numeric values reported back by the AM700 will match the units
set by SCPI, but will not match the displayed values when the front panel
set units and the SCPI set units do not match.
Query: AME[1–4]:SWE:LEV:DEL? returns the numeric value of the minimum level
change that will be used to recognize that a voltage sweep is occuring. The
numeric value reported back matches the units set by SCPI in the UNIT:VOLT
command.
*RST: Set Frequency Min Rise/Fall to 1.
AMEasure[1–4]:SWEep:MODE FREQuency|LEVel
Usage: Sets or queries the type of sweep, frequency or voltage, that will be detected.
Query: AME[1–4]:SWE:MODE? returns FREQUENCY or LEVEL for the setting of
the sweep type detection for the designated measurement.
*RST: Sets sweep detection type to Frequency sweep in all measurements.
AMEasure[1–4]:SWEep:STATe <Boolean>
Parameters: ON|OFF and 1|0
Usage: Sets or queires the state of the AM700 SWEep recognition, ON or OFF
Query: AME[1–4]:SWE:STAT? return 0 for sweep detection disabled or 1 for
enabled in the designated measuement.
*RST: Sets sweep detection to enabled in all measurements.