Teledyne 460L Car Video System User Manual


 
TELEDYNE INSTRUMENTS
Getting Started 460L Instruction Manual
20 05228 Rev B
PRINTED DOCUMENTS ARE UNCONTROLLED DCN 5164
The Serial Port of the device being used to communicate with the 460L should be
configured as follows:
Table 3-6 Serial Port Configuration
PARAMETER VALUE
Baud Rate 9600 bps
Data Bits 8
Stop Bits 1
Parity None
3. Find the address of the 460L.
This can be accomplished by repeating the O3 command (see Section 7.3.5) for each
possible ad
dress (0-9) until the M4
60M responds:
EXAMPLE: 0O3<CR>
1O3<CR>
2O3<CR>
9O3<<CR>
When the instrument responds record the first number in the response line
This is the address of the analyzer.
4. To determine if the M460L is operating correctly type:
1TLIST<CR>
This analyzer will respond with a list of the current values of its test functions. See Section
7.3.8 for a list of the nominal values for these functions
5. C
heck the analyzer’s operation variables (VARS). These describe certain application
specif
ic conditions such as units of measure, measurement ranges, etc.
To check the current state of the VARS, type:
1VLIST<CR>
This analyzer will respond with a list of the current values of its VARS. See Section 7.3.10
for a l
ist of the nomi
nal values for these functions
6. If you need to change one of the VARS settings, see the VSET command (see Table 7-3 of
Section 7.3.11)
7. After viewing the data
you can return the analyzer to normal operation.
After changing the value of any of the VARS return the analyzer to normal operation.
The simplest way to do this is to turn the analyzer off and back on. User Notes:
USER NOTES: