Serial Communications 460L Instruction Manual
38 05228 Rev B
PRINTED DOCUMENTS ARE UNCONTROLLED DCN 5164
7.3.10. VLIST
SYNTAX
<address>VLIST<CR>
DESCRIPTION
Returns a formatted, easy to read list of an internal configuration variables (VAR’s) and their
current values.
DATA PARAMETERS SENT
None
RESPONSE
VAR List (See Below)
EXAMPLE
Command:
1VLIST<CR>
Response:
#0 analog_range = 1000.0
#1 alarm_enable = 1.0
#2 alarm_mode = 0.0
#3 carrier_weight =32.0
#4 comm_mode = 0.0
#5 iir_filt = 0.25
#6 conc_units = 2.0
#7 hi_al_level = 100.0
#8 hihi_al_level = 300.0
NOTES
No checksum is sent in the response to the VLIST command.
The 460L does not use the comm_mode VAR, so its value will always be 0.
hi_al_level and hihi_al_level are returned in the units of measure for which the monitor is set.