5-6
Chapter 5, FRONT PANEL OPERATION
OFF
Output expansion is turned off.
X ONLY
A ×10 output expansion is applied to the X output only.
Y ONLY
A ×10 output expansion is applied to the Y output only.
X & Y
A ×10 output expansion is applied to both the X and Y outputs.
CH1 OUTPUTS CH2
This control, shown on the right-hand LCD, allows the two rear panel analog outputs
CH1 and CH2 to be connected to the required instrument outputs. The left-hand
ADJUST keys are used to select the output provided at the CH1 connector and the
right-hand ones select that at the CH2 connector. Each output may be set to one of
the following settings:-
X %
When set to X % the corresponding rear panel CH1/CH2 connector will output a
voltage related to the X %fs front panel display as follows:-
X %fs CH1/2 Voltage
+120 12.0 V
+100 10.0 V
0 0.0 V
-100 -10.0 V
-120 -12.0 V
Y %
When set to Y % the corresponding rear panel CH1/CH2 connector will output a
voltage related to the Y %fs front panel display as follows:-
Y %fs CH1/2 Voltage
+120 12.0 V
+100 10.0 V
0 0.0 V
-100 -10.0 V
-120 -12.0 V
MAG %
When set to MAG % the corresponding rear panel CH1/CH2 connector will
output a voltage related to the MAG %fs front panel display as follows:-