Comments A one (1) returned in response to the query indicates an overload condition exists.
A zero (0) indicates there is not an overload condition.
Coupling group: none
*RST Condition: none
Example Determining if an overload condition exists
OUTP2:OVER? determine if overload condition exists
[:STATe]
OUTPut2[:STATe] <mode> closes or opens the output relay of the “Main Output”
BNC to enable or disable the analog output. When disabled, the output appears as
an open circuit.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
mode boolean OFF|0|ON|1 none
Comments
Executable when initiated: Yes
Coupling group: none
*RST Condition: OUTPut2:STATe ON
The output relay will open automatically if an output overload is detected.
Example Disabling the output
OUTP2 OFF Disable output
[:STATe]:ACTual?
OUTPut2[:STATe]:ACTual? determines if the amplifier’s ’Main Output’ BNC is
enabled or has been disabled due to an overload condition. The determination is
done by reading bit 8 of the amplifier’s Status register (Appendix C).
This command requires Agilent E1445A firmware revision A.02.00 or greater.
OUTPut2 [:STATe]
OUTPut2 Subsystem E1446/E1445 Command Reference 3-13